Product Description
Atomic emission spectrometer Inductive Coupled Plasma Emission Spectrometer Metal Mineral Analysis Machine
1.Technical parameters
solid state power supply |
Working frequency |
27.12MHz |
frequency stability |
<0.05% |
Matching method |
automatic matching |
Output power |
800W ~ 1600W, continuously adjustable, power efficiency greater than 65% |
Output power stability |
≤0.05% |
The output working coil has an inner diameter of 25mm and 3 turns, and is equipped with three concentric quartz torches with an outer diameter of 20mm. |
The outer diameter of the coaxial sprayer is 6mm; the outer diameter of the double-cylinder mist is 35mm |
scanning beam splitter |
light path |
Czerny turner |
focal length |
1000mm |
Grating specifications |
Ion etching holographic grating, line density 3600L/mm or 2400L/mm; Marking area (80×110)mm |
resolution |
0.008nm (3600 line grating) ≤0.015nm (2400 line grating) |
Scan wavelength range |
3600 line grating: (190~500) nm; 2400 line grating: (190~800) nm |
Electronic measurement and control circuits
Photomultiplier tube specifications: R212/R928
Photomultiplier tube negative high voltage: (-50~-1000) V
Photomultiplier tube current measurement range: (10¹²~10')A
Signal acquisition: VF conversion
computer system
Host: Lenovo brand computer
Monitor: 17-inch LCD monitor
Printer: Canon inkjet printer
Machine technical indicators
Scanning wavelength range: 190-460nm (4320L/mm grating), 190nm~500nm (3600L/mm grating), 190nm~ 800nm (2400L/mm grating);
Measuring range
a. Liquid content: 0.01ppm-several thousand ppm
b. Solid content: 0.001%------70%
Repeatability: (i.e. short-term stability) relative standard deviation RSD≤1.5%;
Stability: Relative standard deviation RSD≤2%;
Test speed: 5-8 elements/minute
Instrument features
More than 70 measurable elements
The analysis speed is fast, and 5-8 elements can be measured in one minute.
Simultaneous analysis of multiple elements, customers can freely choose the number of elements and arrange the measurement sequence
The detection limit is low, reaching ppb level, Ba even reaches 0.7ppb
The linear dynamic range is wide, up to 6 orders of magnitude, and high and low contents can be measured simultaneously
The analysis cost is low, one bottle of argon can be used for 8 hours
Total reflection achromatic optical system
Using concave mirrors instead of convex lenses as optical focusing elements solves the chromatic aberration problem caused by different focus points of different elements and improves the efficiency of the optical system.
Wide wavelength range
Choose 4320 engraved lines grating, the wavelength range can reach 190nm-460nm
Choose 3600 engraved lines grating, the wavelength range can reach 190nm-500nm
Choose 2400 engraved line gratings, the wavelength range can reach 190nm-800nm
Fully automated design
It has a high degree of automation. Except for the power switch, all functions of the entire instrument are controlled by a computer, which is reliable, safe and convenient.
Gas flow automatic control
In the sampling system, the carrier gas, plasma gas, and auxiliary gas are all controlled by advanced mass flow controllers (MFC), which have the advantages of continuously adjustable flow and stable flow, ensuring the stability of the sampling system and ensuring the stability of the light source. Provides strong protection.
Advanced sampling system
The sampling system is efficient and stable, and can be equipped with various atomizers and spray chambers at home and abroad. It can be equipped with high-salt atomizers, hydrofluoric acid-resistant atomizers, etc. to meet the needs of different customers.
Stable waste liquid removal system
Waste liquid is removed through a peristaltic pump to ensure that the sample injection volume is consistent with the waste liquid removal speed. Customers can adjust the speed appropriately as needed to ensure the stability of the sampling system.
*Stable and advanced all-solid-state RF power supply
The RF power supply used by the instrument is a self-developed all-solid-state RF power supply. It has many advantages such as small size, high efficiency, stable output power, and various protection functions, which further improves the stability and safety of the instrument.
*Fast and accurate fully automatic matching function
The load terminal adopts self-developed fully automatic matching technology, which has the advantages of fast matching speed and high accuracy, ensuring output power
Detailed Photos
Our Advantages