Figure 2. The outgoing spectrum of the X-ray tube and the incident spectrum of the characteristic X-ray monochromatized by LSDCC
The correlation coefficient γ of the working curve is greater than 0.999, indicating that the linear error of GA-116 sulfur tester is very small.
| specimen | nominal value(ppm) | Sample cup number one(ppm) | Sample cup number two(ppm) |
| diesel5 | 5 | 4.81 | 4.99 |
| diesel3 | 3 | 2.93 | 3.05 |
| gasoline2 | 2 | 2.12 | 1.82 |
| gasoline10 | 10 | 10.8 | 9.8 |
| gasoline25 | 25 | 24 | 25.3 |
| Sample cup number | Sample number one(ppm) | Sample number two(ppm) | Sample number three(ppm) |
| 1 | 1.19 | 3.21 | 10.22 |
| 2 | 1.08 | 2.92 | 9.79 |
| 3 | 0.85 | 3.05 | 10.09 |
| 4 | 0.93 | 2.88 | 9.73 |
| 5 | 1.05 | 2.86 | 9.93 |
| 6 | 1.03 | 3.16 | 9.90 |
| 7 | 0.98 | 3.26 | 9.95 |
| average value | 1.02 | 3.05 | 9.93 |
| standard deviation | 0.095 | 0.122 | 0.259 |
| RSD | 9.5% | 4.00% | 2.60% |
Main technical index
| X-ray tube | voltage:≤50keV,electricity:≤2mA,power≤50W,target materials:Ag |
| detector | Ultra-thin beryllium window proportional counter |
| LOD(300s) | 0.33ppm |
| measuring range | 1ppm ~ 9.99% |
| up to standard | GB/T 11140,ISO20884,SH/T 0842,ASTMD2622,D7039,D7220 and so on |
| System analysis time | 1~999s,Recommended value: 300s for micro measurement and 60s for major measurement |
| service conditions | environment temperature:5~40ºC,relative humidity:≤85%(30ºC),power supply:220V±20V,50Hz,≤200W |
| Measuring atmosphere | Self-charging system or helium |
| Package Size and weight | 1250mm(W)×690mm(D)×850mm(H),85kg |