Customization: | Available |
---|---|
After-sales Service: | Online |
Warranty: | 1 year |
Still deciding? Get samples of US$ 1000/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Functional items/models | FZS4-4 | FT-100B | FT-100A | FT-100CA | ||
Number of test stations |
1 | 2 | ||||
Test data processing method |
Manual calculation | With printing + automatic calculation | ||||
Use of the measuring cylinder (glass material) for mass determination |
25ml,100ml (Optional10ml) |
25ml,50ml,100ml,250ml(Purchase a 10ml measuring cylinder for testing samples below 25ml). | ||||
density | 0-9.9999g/cm3 | |||||
measurement mode | Measurement Mode | Measurement Mode | ||||
Amplitude | Integer 1-14mm adjustable; | |||||
Vibration frequency | 0~300 times/min (continuously adjustable) | |||||
Vibration time | 0-9999S | none | ||||
Test method | Fixed mass method (standard); Fixed volume method (optional stainless steel combined measuring cylinder parts) | |||||
Fixed volume method combined measuring cylinder |
Equipped with stainless steel measuring cylinders of 25ml; 100ml; (or optional 10ml measuring cylinder for testing samples below 25ml). The combined measuring cylinder consists of an upper sleeve and a calibrated volume measuring cylinder. |
|||||
Reference standard | Specifications of USP, BP, EP Pharmacopoeia; GB/T5162-2006; ISO3953:1993; GB/T31057.2-2018. | |||||
Measuring cylinder requirements |
1.Volume 25±0.5ml, tolerance ±0.25ml; 2. Volume 100±1.0ml, tolerance ±0.5ml; 3. Volume 10±0.2ml, tolerance ±0.10ml Equipped with measuring cylinder cover to solve the problem of powder splashing and overflowing during the compaction process. |
|||||
Repeatability | Error: ≤1% | |||||
Accuracy | Error: ≤1% | |||||
Choice of test method | The fixed mass method is applicable to powder samples that are not easy to stick to the measuring cylinder; the fixed volume method is applicable to fine powder samples that are easy to stick to the measuring cylinder. | |||||
Application areas | Non-metallic powder, metal powder, pharmaceutical, chemical, battery, abrasive, scientific research and teaching and other powder and particle materials | |||||
power supply | AC 220V (50Hz/60Hz), 40W | |||||
Usage Environment | Humidity less than 85%, no condensation, no smoke,no mechanical vibration or electromagnetic interference | |||||
Other Optional Items |
3Q verification document b.Soundproof box c.Moisture meter d.Powder true density meter e.Bulk densitymeter |