Audited by an independent third-party inspection agency
|
Temperature Range
|
RT-600
|
|
DSC Range
|
±1000mw (expandable)
|
|
Temperature Repeatability
|
0.001
|
|
Heating Rate
|
0.1~100/min
|
|
Cooling Rate
|
1~20/min(550~100
|
|
DSC Sensitivity
|
≤0.5uw
|
|
Data Acquisition Frequency
|
Customizable programmable, ≤20 data points collected per second
|
|
Temperature Control Method
|
Intelligent PID temperature control, intelligent memory of temperature control mode, combining the advantages of dynamic and
static algorithms, with self-tuning and self-optimization functions. |
|
Cooling Method
|
Air-cooled
|
|
Flow Meter
|
Mass flow meter, range 0~200ml/min
|
|
Enthalpy Accuracy
|
±0.5%
|
|
Atmosphere Device
|
Built-in dual gas path, compatible with nitrogen, oxygen, air, carbon dioxide, argon, etc.
|
|
Sensor Type
|
Type E
|
|
Sensor Manufacturing Process
|
Heat diffusion process
|
|
Furnace Manufacturing Process
|
Anti-corrosion and anti-oxidation treatment
|
|
Operating Software
|
The experiments can be exported in Excel, PDF, and other formats; the software includes an automatic analysis function, capable of
analyzing specific heat capacity, enthalpy, melting point, glass transition temperature, degree of cure, crystallization, oxidation induction period, etc.; the software has a database of standard test methods; and the software can simultaneously compare more than 10 sets of data. |
|
Program Settings
|
≥50 segments, expandable.
|
|
Comparison Function
|
Standard sample comparison can be performed, with a reasonable error value set; if the error value is exceeded, the comparison
will indicate non-compliance. |